-22-
Instruction Manual
9.
Disconnect the
"VENOUS OUTLET"
line from the reprocessing connector attached to the dialyzer
venous blood port.
Remove the reprocessing connector and place it in fresh 1% RENALIN 100 Solution or
full strength Actril Cold Sterilant for a minimum of thirty (30) minutes.
10.
Place a disinfected blood port cap
on the dialyzer venous blood port
.
H.
Suggested Cleaning of the Dialyzer Case
1.
Wipe the entire capped, reprocessed dialyzer
with a disposable wipe soaked in fresh 1% RENALIN
®
100
solution or full strength ACTRIL
®
Cold Sterilant or place the entire dialyzer in a container of 1% RENALIN
100 solution or full strength ACTRIL Cold Sterilant.
2.
Discard the disposable wipe
.
I.
Suggested Manual Dialyzer Identification and Documentation Procedures
.
IMPORTANT NOTE –
IT IS THE RESPONSIBILITY OF THE PRESCRIBING PHYSICIAN TO DETERMINE
WHAT INFORMATION IS RECORDED AND THE MANNER IN WHICH THE RECORDS ARE MAINTAINED.
1. The use of indelible, waterproof, or permanent ink or other indelible marking systems to label the dialyzer is
recommended when the RENATRON
®
II 100 Series station is operated without RENALOG
®
RM. In the
automatic mode, the information listed below is recorded automatically or entered manually into the RENALOG
RM computer database and an appropriate label is automatically provided for the dialyzer. Additional labels
for use with RENALOG RM are available from MEDIVATORS (PN 78398-179).
2. When reprocessing and not using the RENALOG RM, it is recommended that the following information be
recorded legibly on each dialyzer.
a. Patient’s Identification (required).
b. Dialyzer use number (required).
c. Date the dialyzer was last reprocessed (required).
d. Pressure leak test results (optional).
e. Measured volume of blood compartment test (optional).
f. The initials of the operator who reprocessed the dialyzer (optional).
g. Any other information required by the prescribing physician.
3. It is recommended that a label with the above information be used on every reprocessed dialyzer. A label meeting
these requirements is provided when using the automatic (bar code scanning) features of the RENATRON II
100 Series system.