-28-
Instruction Manual
3. Place the uptake tube into a container of purified water (one gallon or more).
4. Connect external hoses in sanitize configuration.
5. Turn on the RENATRON
®
II 100 Series Station. Press the
"MUTE ALARM"
and
"RESET"
switches to
change to the
"HF"
mode. Press the following switches in the following order:
"MUTE ALARM"
"RESET"
"DISPLAY"
The Volume display window will read
"409"
and the
"PROGRAM STEP"
display window will read "SC".
Press the
"START SANITIZE"
switch to begin the cycle.
6. Allow the machine to operate to completion of "PROGRAM STEP"
75
and press the
"OFF"
switch. (No
problems will occur if machine operates past step
75
before turning off).
7. Press the
"ON"
switch and repeat procedure steps
5
and
6
above. This action clears RENALIN
®
100 from the
uptake hose and replaces it with purified water.
8. Remove the uptake tube from the purified water and allow excess water to drain out of the uptake tube.
9. Place the uptake tube into container of Formula 409.
Do not use a substitute chemical as it may damage the
materials in the machine
.
10. Press the
"ON"
switch and repeat procedure steps
5
and
6
above. This action clears the water from the uptake
hose and replaces it with Formula 409.
11. Turn on the RENATRON II 100 Series Station. Press the
"MUTE ALARM"
and
"RESET"
switches to
change to the "HF" mode. Press the following switches in the following order:
"MUTE ALARM"
"RESET"
"DISPLAY"
The Volume display window will read "409" and the PROGRAM STEP display window will read "SC". Press
the
"START SANITIZE"
switch to begin the cycle.
12. Allow the machine to continue operating to the
"SANITIZE COMPLETE
" message in
PROGRAM STEP
83
.
13. Press the
"OFF"
switch and turn the water supply to the machine off. Leave the machine overnight.
Wait at
least 8 hours
.
14. Remove the uptake tube from the container of Formula 409 and make sure any excess Formula 409 drains from
tube back into container. Place the uptake tube in a container of purified water (one gallon or more). Turn the
water supply on.
15. Turn on the RENATRON II 100 Series Station. Press the
"MUTE ALARM"
and
"RESET"
switches to
change to the "HF" mode. Press the following switches in the following order:
"MUTE ALARM"
"RESET"
"DISPLAY"
The Volume display window will read "409" and the PROGRAM STEP display window will read "SC". Press
the
"START SANITIZE"
switch to begin the cycle.
16. Allow the machine to operate to completion of "PROGRAM STEP"
75
and press the
"OFF"
switch. (No
problems will occur if machine operates past step
75
before turning off.) This action clears the Formula 409
from the uptake hose and replaces it with water.
17. Turn on the RENATRON II 100 Series Station. Press the
"MUTE ALARM"
and
"RESET"
switches to
change to the "HF" mode. Press the following switches in the following order:
"MUTE ALARM"
"RESET"
"DISPLAY"
The Volume display window will read "409" and the PROGRAM STEP display window will read "SC". Press
the
"START SANITIZE"
switch to begin the cycle.
18. Allow the machine to operate to PROGRAM STEP
83
, and when machine completes step
83
, press the
"START
PROCESS"
switch. The machine will then enter step
84
and, after a ten minute hold period in step
84
, will
proceed with flushing and draining. The flushing and draining will continue until machine enters PROGRAM
STEP
97
and the
"SANITIZE COMPLETE"
message reappears.
19. Press the
"OFF"
switch.
20. Remove the uptake tube from the container of purified water and allow excess water to drain from uptake tube.
21. Place the uptake tube in a container of unexpired RENALIN 100 solution.