8.2 Before Calling a Service Person
– 49 –
D
u
ri
ng
me
a
s
u
rem
en
t
*
Reliability has been determined
to be low after multiple measure-
ment data have been analyzed.
The cause of the low reliability
may be:
(1) Eye has moved.
(2) Reflection rate of examinee’s
eye is low.
Measure again.
D
u
ri
ng
pr
in
ti
n
g
PRINT CANCELLED
Printing has been cancelled by
pressing one of the switches
during printing.
Perform printing again as
required.
NO PRINT DATA
There is no measurement data.
Print after measurement.
PRINT: OFF
“PRINT” is set to “OFF” in SET
mode.
Set “PRINT” to “ON” as required.
PRINT: OFF / TRANS: OFF
Both “PRINT” and “TRANS” are
set to “OFF” in SET mode.
Set them to “ON” as required.
PRINTER LEVER IS DOWN
Printer lever is lowered.
Raise the lever.
NO PRINT PAPER
There is no printing paper.
Load role of printing paper.
PRINTER TOO HOT
The printer head is too hot.
Wait for about 5 minutes and try
again.
PRINTER TOO COLD
The printer head is too cold.
D
u
ring
tran
sf
er
TRANSMITTING
Data is being transmitted.
-----
TRANS COMPLETED
Data transfer has been com-
pleted.
-----
NO TRANS DATA
There is no measurement data.
Transfer after performing mea-
surement.
TRANSMISSION CANCELLED
Transmission has been can-
celled by pressing one of the
switches during transfer.
Transfer again as required.
RECEIVED BREAK CHARACTER
The instrument has received a
break character from the exter-
nal equipment.
Transfer again as required.
COMMUNICATION ERROR
(401)
The connected equipment is not
ready yet or not connected.
Prepare the external equipment.
COMMUNICATION ERROR
(402) - (410)
Format set in SET mode and
that of the external equipment
do not match.
Change the settings as required.
Message / Display
Cause
Remedy
Summary of Contents for TX-F