7.1 Settings for Measurement (Page: 1/3)
– 33 –
7.1 Settings for Measurement (Page: 1/3)
7.1.1 Basic Operation
How to Select the Item
Turn the roller to select the item to be changed.
The selected item will be displayed in reverse image.
How to Change the Setting
Press either
FIXATION
switch,
LIMITER
switch,
START
switch or
switch.
The selected setting will be highlighted.
NOTE: Some settings must be changed in different ways. The procedure for changing
those settings will be indicated under the explanation of each item.
7.1.2 Items
RANGE
Measurement range
“
30
”:
Used for measuring the range 0 - 30 mmHg.
If the result is over 30 mmHg, the range automatically
changes to “60”. This is recommended to be used for
group examinations and normal screening.
“
60
”:
Used for measuring the range of 25 - 60 mmHg. (The
air puff is stronger.)
If the result is under 25 mmHg, the range automati-
cally changes to “30”. This is recommended to be used
for examination of glaucoma, etc.
PAGE
END
PAGE
S E T M O D E
R A N G E
;
3 0
6 0
C O U N T
:
O N
O F F
O R D E R
W A R N I N G I O P
:
O F F
2 5
W A R N I N G I M G
:
O N
O F F
:
S E QU
R E L I
DAT E
:
2 5 / S E P / 2 0 0 2
1 4 : 5 0
E R RO R :
O N
O F F
[ N o . ] :
= 0 1 2 3 4 5
R E S E T
[ O D R ]:
Y M D
M DY
D M Y
PAG E : 1 / 3
U N D E R
A B O V E
: