![Canon OMNERA 400T Operator'S Manual Download Page 97](http://html.mh-extra.com/html/canon/omnera-400t/omnera-400t_operators-manual_61479097.webp)
Operator´s Manual
0180-095-418-EN Rev. 6.1
OMNERA®
2020-09-18
Page 89
Wall stand synchronization
WARNING!
Before performing any wall stand tracking, assure that the wall stand indication light
is lit and thereby, that the wall stand is activated.
At wall stand synchronization, the collimator reticle shall be aligned with the detector cross.
When performing fast or long movements of the wall stand detector, it may occur that the col-
limator does not synchronize with the wall stand detector. The automatic wall stand tracking
may not make it all the way and the indication light will start flashing.
In this case, activate the synchronization button at the wall stand, see Fig. 4-8
Fig. 4-8 Wall stand synchronization button
Then the tracking will carry out the full movement and synchronize. The indication light will
shine permanently.