OMNERA®
0180-095-418-EN Rev. 6.1
Operator´s Manual
Page 40
2020-09-18
Guidance and manufacturer’s declaration - electromagnetic immunity
The System is intended for use in the electromagnetic environment specified below. The customer or
the user of the System should assure that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the System, including cables, than the
recommended separation distance, calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance;
Conducted
RF IEC
61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80
MHz
d = 1.2 √p
Radiated RF
IEC 61000-4-
3
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHz to
2.5 GHz
d = 1.2 √p
80 MHz to 800 MHz
d = 2.3 √p
800 MHz to 2.5 GHz
where
p
is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the recom-
mended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be range.
b
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre-
dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans-
mitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the System is used exceeds the applicable RF compliance level above, the System
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as relocating the System.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Operator s Manual 0180 095 418 EN Rev 6 1 ...
Page 3: ...Rev Date New release 6 1 2020 09 ...
Page 4: ......
Page 8: ...Contents OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page iv 2020 09 18 ...
Page 148: ...OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page 140 2020 09 18 Waste disposal ...
Page 165: ...Operator s Manual 0180 095 418 EN Rev 6 1 OMNERA 2020 09 18 Page 157 Appendix B Annual Checks ...
Page 166: ......