Wall stand tracking technique
General description
The
Wall stand tracking
technique is intended for examinations against a wall stand.
In this technique the tube holder will track the movements of the wall stand to assist the oper-
ator to always keep the correct position to the detector.
Wall stand tracking
The tube holder can track the wall stand in two different positions depending if the detector is
placed vertically or horizontally.
The system decides which way depending on the angle of the X-ray tube.
If the angle is between -45 and +45 degrees the detector is assumed to be horizontally
placed and the default tracking distance for a horizontally placed detector is chosen.
The SID is shown on the display.
If the angle is outside -45 to +45 degrees the system assumes that the detector is placed ver-
tically and the vertical default tracking distance is chosen.
No SID or height is shown on the display. The default tracking distances are set during instal-
lation of the system.
Note!
In Wall stand tracking technique, the exposure is blocked whenever a table workstation is
chosen.
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Operator s Manual 0180 095 418 EN Rev 6 1 ...
Page 3: ...Rev Date New release 6 1 2020 09 ...
Page 4: ......
Page 8: ...Contents OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page iv 2020 09 18 ...
Page 148: ...OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page 140 2020 09 18 Waste disposal ...
Page 165: ...Operator s Manual 0180 095 418 EN Rev 6 1 OMNERA 2020 09 18 Page 157 Appendix B Annual Checks ...
Page 166: ......