3.15.3.5 Operating the automatic collimator, wall stand
For further information of how to operate the wallstand automatic collimator, see 3.15.3.3
erating the automatic collimator, table
Top and bottom centering
The collimator light field can be top or bottom aligned instead of centered against the
detector.
For top centering this means that the upper border of the collimator light field is aligned with
the top of the detector.
For bottom centering the collimator light field is aligned with the bottom of the detector. The
stand will automatically request synchronization to keep the alignment of the top or bottom of
the detector when the collimator light field is increased or decreased. The functionality of top
and bottom centering is available on a vertical receptor on the wallstand.
With top centring active is the light field
always as high as it can be regarding size
and position for the image receptor .
Fig. 3-37 Top and bottom centering
With the top centering active, the light field always is as high as it can be regarding size and
position for the image receptor
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Operator s Manual 0180 095 418 EN Rev 6 1 ...
Page 3: ...Rev Date New release 6 1 2020 09 ...
Page 4: ......
Page 8: ...Contents OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page iv 2020 09 18 ...
Page 148: ...OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page 140 2020 09 18 Waste disposal ...
Page 165: ...Operator s Manual 0180 095 418 EN Rev 6 1 OMNERA 2020 09 18 Page 157 Appendix B Annual Checks ...
Page 166: ......