7.2 Monthly checks
7.2.1 Checklist
7.2.1.1 Ceiling suspended X-ray tube support
1. Move the ceiling suspended X-ray tube support
manually to all positions in X, Y and Z direction and
make sure it runs smoothly and sounds OK.
2. Check the emergency stop. By activating the emer-
gency stop all motorized movements are inhibited.
3. Choose table position and make sure tracking is
activated.
Measure between the X-ray tube focal spot and the
active image receptor surface of the detector
holder.
The measured SID shall correspond with the dis-
played SID.
Move the ceiling suspended X-ray tube support in X
or Y direction and measure between the X-ray tube
focal spot and the active detector surface of the de-
tector holder again.
The SID is allowed to differ ±1%.
4. Check that the SID, shown on the display of both
the Image system and the collimator, correspond
with the measured SID/FFD.
5. Check the hoses for damage.
6. Check all outer cabling for damage.
7. Clean all outer surfaces, except for the lubricated
column segments. See Instruction for use for clean-
ing instructions.
8. Power up the ceiling stand and check all functions.
9. Move the ceiling stand around and observe any
irregularities.
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Operator s Manual 0180 095 418 EN Rev 6 1 ...
Page 3: ...Rev Date New release 6 1 2020 09 ...
Page 4: ......
Page 8: ...Contents OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page iv 2020 09 18 ...
Page 148: ...OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page 140 2020 09 18 Waste disposal ...
Page 165: ...Operator s Manual 0180 095 418 EN Rev 6 1 OMNERA 2020 09 18 Page 157 Appendix B Annual Checks ...
Page 166: ......