2.10.3.3 Weight Restrictions Table 0055
The following figures show the maximum load at different positions of the table.
When the table top is centered over the table frame, the maximum load of a patient either ly-
ing or sitting is 300 kg/, 611 lb, see Fig. 2-10
Maximum patient load, patient centered
002
277_0
072
MAX. 300 kg / 611 lb
Fig. 2-10 Maximum patient load, patient centered
When the table top is positioned outside the table frame, the maximum load of at patient lying
on the table top is 200 kg/ 440 lb and the maximum load of at patient sitting on the table top
is 150 kg / 330 lb.
4097
B
C
C
Fig. 2-11 Maximum patient load position B – C
Table 2-2
Pos.
Model 0055
B
200 kg / 440 lb
C
150 kg / 330 lb
The table frame is marked on the upper side with the maximum weight when positioning in
outer positions, see Fig. 2-12