3.15.3.3 Operating the automatic collimator, table
Startup Mode
At startup of the system, the collimator is defined to
Automatic mode
, light field to
Maximum
,
SID to
110 cm
and filter to the first defined.
Finding the right position
The programmed SID value used for calculating the size of the light field is shown in the dis-
play of the automatic collimator. The correct position of the X-ray tube support is reached
when the true SID value corresponds with the programmed SID value shown on the auto-
matic collimator display.
The SID value is shown when wallstand or table tracking is activated, the detector/detector
holder is in a horizontal position and the tube alpha angle is between –45 and +45 degrees.
When tracking is not activated and the detector holder is not horizontal, a measuring tape
may be used to determine the SID.
When performing examinations at the wallstand with the detector holder in a vertical position
the positioning indexes at the ceiling rails can be used in order to fast find the correct
position.
Automatic collimator light
When the tube stand is tracking against the wallstand or when the table top is released, the
collimator light will automatically be turned on.
Changing working mode
The collimator can be operated in either
Automatic
or
Manual mode
. The purpose of the
Manual mode
is to be able to adjust the collimator light field outside the detector.
The mode can easily be changed on the collimator (button 6) or at the collimator control han-
dle (button H).
Automatic mode
The maximum light field size is restricted to the detector size.
Detector size
When tracking Table/WS is activated;
Adjust the collimator light field to the detector size by pushing and holding the
M-button
on
the collimator control handle (button H) or on the collimator (button 6) for approximately 2
seconds.
SID
Changing SID
The SID used for calculating the size of the light field can be changed manually with button 9
on the collimator.
The new SID value will then be used for calculating the field size instead of the pre-pro-
grammed SID value.
Note!
The System steps between pre-set values: 100, 115, 150, 180, 200.
Pre-programmed SID values
If the SID values for each APR are pre-programmed at the Canon NE user interface, this will
override the
collimator default value
.
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Operator s Manual 0180 095 418 EN Rev 6 1 ...
Page 3: ...Rev Date New release 6 1 2020 09 ...
Page 4: ......
Page 8: ...Contents OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page iv 2020 09 18 ...
Page 148: ...OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page 140 2020 09 18 Waste disposal ...
Page 165: ...Operator s Manual 0180 095 418 EN Rev 6 1 OMNERA 2020 09 18 Page 157 Appendix B Annual Checks ...
Page 166: ......