Appendix — Technical description —
82
This link action establishes infrared link communication between the infrared
communication section of the detector and the ready indicator connected to the image-
capture computer.
5.2.1
Communication timeout
3 sec (The period during which the detector tries infrared communication after the
detector’s POWER switch is pushed down.)
5.2.2
Communication environment
Communication range:
Within 30 cm
Environment conditions: Cannot be used in an environment in which infrared radiation is
blocked.
Use items selected by your sales representative for the following units:
• Wireless LAN access point
• LAN cable
• Micro USB cable (used to connect the ready indicator)
• Image-capture computer
Service technicians who have undergone training for servicing this product, including
specifications, installation procedures and parts replacement procedures. Limited to
those with permission from the department that hosts training.
Please never perform the installation, repair, maintenance check except an above service
engineer, for the product works safety and enough performance.
Work-related precautions
Bear the following points in mind when installing the system.
(1)
Carry out the work while ensuring that none of the portable units will be dropped or
subjected to impact.
(2)
Ensure that the installation location satisfies the following criteria:
1) A location where the product will not be splashed with water
2) A location where the air does not contain moisture, dust, salt, sulfur or other such
substances and where the product will not be adversely affected by poor ventilation,
the air pressure level, the temperature and other such factors
3) A location which is not tilted or subject to vibration or impact (this includes the
locations when the product is conveyed)
4) A location where chemical products are not stored and where no gases are given off
5.2 Infrared LINK communication specifications
6 Other items to be installed
7 Installation, repair and maintenance check
7.1 License for service engineer
7.2 Installation procedures
000_DIA-2019-019-E03_AX-C4343W.book Page 82 Friday, November 27, 2020 12:38 PM
Summary of Contents for AX-C4343W
Page 6: ...Conventions 6 000_DIA 2019 019 E03_AX C4343W book Page 6 Friday November 27 2020 12 38 PM ...
Page 51: ...7 Maintenance 000_DIA 2019 019 E03_AX C4343W book Page 51 Friday November 27 2020 12 38 PM ...
Page 53: ...8 Specifications 000_DIA 2019 019 E03_AX C4343W book Page 53 Friday November 27 2020 12 38 PM ...
Page 85: ...000_DIA 2019 019 E03_AX C4343W book Page 85 Friday November 27 2020 12 38 PM ...