Guidance and manufacturer’s declaration for EMD
70
The AX-C4343W is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the AX-C4343W can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the AX-C4343W as
recommended below, according to the maximum output power of the communications
equipment.
Table 4
Recommended separation distances
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
d=1.2
80 MHz to 800 MHz
d=1.2
800 MHz to 2.7 GHz
d=2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
P
P
P
000_DIA-2019-019-E03_AX-C4343W.book Page 70 Friday, November 27, 2020 12:38 PM
Summary of Contents for AX-C4343W
Page 6: ...Conventions 6 000_DIA 2019 019 E03_AX C4343W book Page 6 Friday November 27 2020 12 38 PM ...
Page 51: ...7 Maintenance 000_DIA 2019 019 E03_AX C4343W book Page 51 Friday November 27 2020 12 38 PM ...
Page 53: ...8 Specifications 000_DIA 2019 019 E03_AX C4343W book Page 53 Friday November 27 2020 12 38 PM ...
Page 85: ...000_DIA 2019 019 E03_AX C4343W book Page 85 Friday November 27 2020 12 38 PM ...