36
CT-2000 Instruction Manual
Troubleshooting
A successful CT-2000 self-test is an indication that there are no detectable errors and checked
components are functional. There are certain conditions, however, that cannot be verified by the
automatic tests. These include:
1.
Viscosity of the bath liquid
2.
Temperature probe installation
3.
Motor-stirrer functionality
4.
Temperature probe functionality
5.
Heater operation
Problems with the first three items listed above (bath liquid viscosity, temperature probe
installation and motor-stirrer functionality) may cause incorrect temperature stability. If items 4 or 5
fail (temperature probe functionality and/or heater operation), the bath may be non-operational.
Consult Table 4 for assistance in troubleshooting your CT-2000. If the following table fails to resolve
the difficulties, call your local CANNON representative for assistance.
Table 4: Troubleshooting Causes and Solutions
Problem
Possible Cause/Solution
Bath does not appear to have power
•
Power cable is not connected to outlet or connected
to rear electrical drawer.
•
Power cord is unplugged. Attach cord to appropriate
outlet.
•
Power is out on MAINS. Restore power.
Bath illumination not functioning
•
Separate switch for LIGHT must be on.
•
Florescent lamps may be defective. Replace lamps.
•
Lamps may be out of socket. Tighten lamps.
•
Lamp ballast may be defective. Replace lamp ballast.
•
Ballast power supply may not be working properly.
Replace ballast power supply.
Bath liquid not agitated
•
Stirring motor not functioning. Check connection of