CANNON Instrument Company | Calibrating the CT-2000
25
Calibrating the CT-2000
The CT-2000 has the ability to be calibrated to a temperature reference standard for any desired
temperature, Celsius or Fahrenheit. The controls for entering the target (desired) temperature and
the calibration at this temperature are convenient and largely intuitive. The instrument stores
calibration corrections for every temperature integer.
Procedure
Use the
CAL TEMP
keypad option to calibrate the CT-2000 to your ASTM thermometer for a given
temperature.
Note:
Before calibrating the CT-2000, make sure the bath has stabilized at the
desired temperature.
1.
Set the CT-2000 to the desired test temperature.
2.
When the CT-2000 indicates that the instrument is controlling at the desired temperature, place
the thermometer in the bath and wait for the temperature reading to stabilize (at least 10
minutes).
3.
Press the
CAL TEMP
key and enter the thermometer reading up to three decimal places,
subject to the following parameters:
a.
A decimal point may be included as an entry.
b.
The entered temperature must be within the operational temperature range of the bath. If
not, a
CALIBRATION ERROR
is displayed on the LCD screen.
c.
The entered actual temperature must not cause a correction in temperature greater than
± 2.5 °C.
Note:
If an error is made when entering the temperature, press the
CAL TEMP
button again to cancel the operation, exit the calibration screen and reurn to
normal CT-2000 operation.
4.
When data entry is complete, press
ENTER
to store the calibration data.