12
CT-2000 Instruction Manual
11.
Place the wider (tempered) piece of glass in the front slot furthest away from the jar. The middle
slot is left empty to create a vapor barrier.
12.
Replace the rear top cover. Align the four holes, then insert and tighten the previously removed
screws.
13.
Replace the front top cover. Line up the four holes, then insert and tighten the previously
removed screws.
14.
Ensure that the gasket forms a tight seal with the top covers of the bath:
a.
Loosen the IEC lock screw securing the AC power cord to the rear panel of the electronics
drawer.
b.
Unplug all three cables (AC power, rectangular Cinch connector and round Amp connector)
from the rear of the electronics drawer.
c.
Pull out the electronics drawer using the handles installed on the front of the unit. Press
down or pull up on the plastic release bars on either side of the drawer track to release the
drawer, then pull the drawer completely free of the unit and set it aside.
d.
When the drawer is removed, locate the four ¼-20 set screws visible at the top of the
drawer opening underneath the bath.
e.
Turn the set screws clockwise with the included Allen wrench until the top of the jar forms a
tight seal with the covers. Tighten the set screws uniformly so the jar remains level.
f.
Run the AC cord through the rear panel opening.
g.
Replace the drawer in the slide tracks and push the drawer back into its opening. Insert the
two power plugs into the rear of the drawer assembly.
15.
Remove the motor-stirrer from its box. Remove the two screws on the top heater housing and
lift off the housing. Refer to Figure 12.