CANNON Instrument Company | Communication Options
29
Communication Options
Several communication options between the CT-2000 and your computer may be configured from
the
MENU
keypad option.
Change Port Speed
The
Change Port Speed
option provides a speed setting for data transfer between the CT-2000 and
a computer via the RS-232C connection. Select the baud rate (the default for most software is 9600)
in bps (bits per second) from one of the following options:
1.
1200
2.
2400
3.
8600
4.
19.2k
5.
38.4k
Keypad Sequence
MENU
>
3
>
ENTER
>
1
>
ENTER
>
Input speed choice
(1 – 5)
>
ENTER
MODE
The
MODE
keypad option permits RS-232C or RS-485 communication options.
The CT-2000 is equipped with a full-duplex RS-232C point-to-point interface as well as a half-duplex
RS-485 network interface. Commands and Queries may be sent to the CT-2000 via either interface.
Responses and Reports may be received by the master controlling computer.
This interface is configured as 8 data bits with no parity and one stop bit.
The mode of operation, polled (multi-drop) or point-to-point, can only be configured via the front
panel menu.
Procedure
To select the
MODE
option, press
2
from the
COM Setup
menu.