GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision E
CONFIDENTIAL
Page 160
The following test is not required for “LE” systems.
10mm, 6.0J/cm2, 0.50ms
Expected Energy = 4.7J
10mm, 12.0J/cm2, 0.50ms
Expected Energy = 9.4J
Pulse # OPHIR Energy, J Pulse # OPHIR Energy, J
1
1
2
2
3
3
Avg
Avg
%Diff
%Diff
27.
Circuit Calibration is complete; the laser can safely be used.