FINAL VERIFICATION OF USER CALIBRATION ENERGY
The final step is to complete user calibrations and to verify the energy is within
specification.
23.
Press the Exit button. This will return the user to the Main user screen.
24.
Complete the following Calibrations using the specified spot size, fluence,
and pulse width from the tables below. After each Calibration, enter Ready
and Pulse 3 times into the Ophir meter. Record each Ophir energy.
25.
Calculate the Average Ophir energy and then the percentage difference
from the expected energy.
26.
Verify the percentage difference of each table is -14%. If this
verification fails, the Circuit CAL needs to be repeated. If it fails more than
once, then contact Candela Customer Service for service.
15mm, 6.0J/cm2, 3ms
Expected Energy = 10.6J
15mm, 44.0J/cm2, 3ms
Expected Energy = 77.7J
Pulse # OPHIR Energy, J Pulse # OPHIR Energy, J
1
1
2
2
3
3
Avg
Avg
%Diff
%Diff