Replace Can
(Only with
DCD Option)
Canister pulse count
reaches zero “0”, or when
air detected at the canister.
Replace DCD canister & reset
canister counter when ever this
message appears.
Purge
(Only with
DCD Option)
Air was detected in the
cryogen lines. Bubble
percentage is outside of
tolerable range (~15%
bubbles)
With handpiece removed from
the calibration port, depress the
purge button until the message
clears.
Exit to Clean
Window
Delivery System
Transmission is low.
Transmission is < 75% -
Delivery system degrading
-
Examine HP Slider window. If
necessary, clean or replace
window.
Candela will make available on request circuit diagrams, component part lists,
descriptions, calibration instructions, or other necessary information which will assist
the customer’s appropriately qualified technical personnel to repair those parts of
equipment which are designated by Candela as repairable.