GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 102
CAUTION!
THE CONTENTS OF THE CRYOGEN CANISTER ARE UNDER
PRESSURE. READ THE MATERIAL SAFETY DATA SHEET
(MSDS) AND THE LABEL ON THE CANISTER BEFORE
HANDLING.
Follow the instructions shipped with each replacement canister.
Additionally, once the canister is installed, the canister count must be reset. If the
newly installed canister is at room temperature or cooler, a 20 - 25 minute WARM -
UP period will be needed prior to use. The WARM -UP screen will appear once the
laser is switched to the READY state if additional warming is required.
A canister warmer is available and may be used to hold canisters and minimize the
warm–up period.
If a major cryogen leak occurs, disconnect the delivery system from the laser or
remove the canister from the laser. Ventilate the room thoroughly and call Service to
correct the problem.
For additional information, refer to the MSDS sheets supplied with each cryogen
canister.
CRYOGEN
CANISTER
REPLACEMENT
MAJOR
CRYOGEN
LEAK
CANISTER
WARMER
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...