5.1
CAL completion HV sample to Ready entry HV smp not
-3% at End of Charge (EOC) OR
HV Reference and HV sample not -5% at End of
Charge
5.2 No EOC signal present within 3 secs after HV setting
5 – HV
TOLERANCE
FAULT
5.3
No EOC signal present after DCD pre-spray already
occurred.
6.2
Expected Head Energy (xHD) is calculated > Max CAL
HD Energy of 100 J.
6.3
Laser failed to CAL to desired Fluence within 20 pulses.
6 – CALIBRATION
FAULT
6.4
CAL required a HV > 1200V for desired Fluence.
7.1
DI temp < 60degC when not in warm-up.
7.2
DI temp > 70degC while in Ready state.
7.3
DI pressure switch does not change when power turned on.
OR
DI pump is not ON or DI pressure switch not actuated.
7 – DI WATER
SYSTEM
FAULT
7.4
DI temp < Thermistor Open Temperature (5) OR
DI temp > Thermistor Shorted Temp (98) while in Ready
state
8.1
DCD pressure < 105 psi while in READY & INT_DCD &
HP w/DCD & spray setting is non-zero.
8.2
DCD Pressure > 135 psi & INT_DCD
8 – DCD SYSTEM
FAULT
8.3
DCD Valve Current was not detected while spraying
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...