GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 83
•
Fully insert the handpiece into the Cal Port.
•
Press the CALIBRATE button on the Display Panel and follow the
instructions p rovided in the calibration message window.
•
Remove handpiece from the Calport when instructed to do so and install
distance
Pressing “X” allows the operator to return to the Main screen in Standby. The
operator can then adjust the laser output parameters as needed before restarting the
calibration procedure.
After successful completion of the calibration procedure, remove the Delivery
System from the Calport, install the Distance Gauge and begin treatments.
During calibration the software performs a delivery system transmission check. If
the delivery system transmission is low, the software will display a message to clean
the HP Slider window. The user can stop the calibration by pressing “YES” and
clean the HP Slider window (see Section 6 for window cleaning procedure) or press
“NO” to continue the calibration. The performance of the delivery system is
improved by using a clean or new HP Slider window. Clean optics will increase the
life of the delivery system.
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...