GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 27
GentleYAG
Laser System Electrical Requirements:
220 V - 230 V~, 50/60 Hz,
3680 VA
Installation Site
Electrical Service Requirements
United States
208 V - 240 V~, 60 Hz, 30 Amp, center-tapped, single phase,
dedicated branch circuit with earth ground conductor.
Worldwide
220 V - 230 V~ (+/-10%) 50/60 Hz, >16 Amp*, single phase,
dedicated branch circuit with earth ground conductor.
*Note:
GentleYAG
Laser System may be suitable for hookup to
a 230 V~, 16 Amp dedicated service. Voltage dips and low line
conditions may cause the mains breaker to trip periodically.
The user should consult a qualified Electrician to verify mains
quality before connecting to a 16 Amp service.
GROUND CONTINUITY TESTS
Your laser system requires a connection to earth ground to reduce the risk of electric
shock. To verify that this safety feature is functioning properly, we recommend that
continuity between the laser chassis and mains plug grounding pin be checked
annually at a minimum, monthly if the laser is moved frequently, or before use if the
line cord and/or power plug has been altered or replaced. If you are unsure of which
TABLE 1-2
ELECTRICAL
REQUIREMENTS
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...