CS616 and CS625 Water Content Reflectometers
In some applications, installation can be improved by using the CS650G
insertion guide tool. The CS650G is inserted into the soil and then removed.
This makes proper installation of the water content reflectometer easier in
dense or rocky soils.
6.3 Wiring
TABLE 6-1. CS616/625 wiring code.
color function datalogger
connection
red
+12 V
+12 V
green
output
SE analog channel
orange enable
control
port
black signal
ground
G
clear shield
(power ground)
G
Both the black ground wire and the clear shield wire must be
connected to datalogger ground.
NOTE
power
Red
output
gnd
Black
output
Green
enable
Orange
drain/power
gnd
Clear
FIGURE 6-1. CS616 Water Content Reflectometer