Section 8. Operation
397
Offset Voltage Compensation
Related Topics
• Auto Self-Calibration — Overview
(p. 92)
• Auto Self-Calibration — Details
(p. 373)
• Auto Self-Calibration — Errors
(p. 515)
• Offset Voltage Compensation
(p. 347)
• Factory Calibration
(p. 89)
• Factory Calibration or Repair Procedure
(p. 500)
Summary
Measurement offset voltages are unavoidable, but can be minimized.
Offset voltages originate with:
• Ground currents
• Seebeck effect
• Residual voltage from a previous measurement
Remedies include:
• Connect power grounds to power ground terminals (G)
• Use input reveral (RevDiff = True) with differential measurements
• Automatic offset compensation for differential measurements when
RevDiff = False
• Automatic offset compensation for single-ended measurements when
MeasOff = False
• Better offset compensation when MeasOff = True
• Excitation reversal (RevEx = True)
• Longer settling times
Voltage offset can be the source of significant error. For example, an offset of 3
μV on a 2500 mV signal causes an error of only 0.00012%, but the same offset on
a 0.25 mV signal causes an error of 1.2%. The primary sources of offset voltage
are ground currents and the Seebeck effect.
Single-ended measurements are susceptible to voltage drop at the ground terminal
caused by return currents from another device that is powered from the CR3000
wiring panel, such as another manufacturer's comms modem, or a sensor that
requires a lot of power. Currents >5 mA are usually undesirable. The error can
be avoided by routing power grounds from these other devices to a power ground
G terminal on the CR3000 wiring panel, rather than using a signal ground (
)
terminal. Ground currents can be caused by the excitation of resistive-bridge
sensors, but these do not usually cause offset error. These currents typically only
flow when a voltage excitation is applied. Return currents associated with
voltage excitation cannot influence other single-ended measurements because the
excitation is usually turned off before the CR3000 moves to the next
measurement. However, if the CRBasic program is written in such a way that an
excitation terminal is enabled during an unrelated measurement of a small voltage,
an offset error may occur.
The Seebeck effect results in small thermally induced voltages across junctions of
dissimilar metals as are common in electronic devices. Differential
measurements are more immune to these than are single-ended measurements
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......