Section 8. Operation
382
Analog Voltage Input Ranges and Options
Range Code
Description
2
Append with C to enable common-mode null / open-input detect (Example:
mV200C)
Input Limits / Common-Mode Range
Related Topics:
• Voltage Measurements — Specifications
• Voltage Measurements — Overview
(p. 68)
• Voltage Measurements — Details
(p. 380)
Note This section contains advanced information not required for normal
operation of the CR3000.
Summary
• Voltage input limits for measurement are ±5 Vdc. Input Limits is the
specification listed in Specifications
(p. 95).
• Common-mode range is not a fixed number. It varies with respect to
the magnitude of the input voltage.
• The CR3000 has features that help mitigate some of the effects of
signals that exceed the Input Limits specification or the common-mode
range.
With reference to the figure PGIA with Input-Signal Decomposition
(p. 383),
the
PGIA processes the voltage difference between V+ and V–. It ignores the
common-mode voltage, or voltages that are common to both inputs. The figure
shows the applied input voltage decomposed into a common-mode voltage (V
cm
)
and the differential-mode component (V
dm
) of a voltage signal. V
cm
is the
average of the voltages on the V+ and V– inputs. So, V
cm
= (V+ + V
–)/2 or the
voltage remaining on the inputs when V
dm
= 0. The total voltage on the V+ and
V–
inputs is given as V+ = V
cm
+ V
dm
/2, and V–
=
V
cm
– V
dm
/2, respectively.
The PGIA ignores or rejects common-mode voltages as long as voltages at V+
and V– are within the Input Limits specification, which for the CR6 is ±5 Vdc
relative to ground. Input voltages wherein V+ or V–, or both, are beyond the ±5
Vdc limit may suffer from undetected measurement errors. The Common-Mode
Range defines the range of common-mode voltages that are not expected to
induce undetected measurement errors. Common-Mode Range is different than
Input Limits when the differential mode voltage in non-negligible. The following
relationship is derived from the PGIA figure as:
Common-
Mode Range = ±5 Vdc
– |V
dm
/2|.
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......