Appendix E. Calculating Measurement Error
E-3
E.5 Example Error Calculation: DGSI Piezometer
52611099
For Durham Geo Slope Indicator piezometer 52611099, when using gauge factors
A
,
B
, and
C
, the formula for pressure,
P
, derived from frequency is given as:
P = Af
2
+ Bf + C
The resulting noise (effective resolution) formula is then:
∆P = (2Af + B)N
or
∆P = 2AfN + BN