Caution
•
Unauthorized disassembly and/or repair of the analyzer by other than an
authorized Cadex Service Center will void the warranty.
•
Observe anti-static precautions before and during service. The circuitry used
in the analyzer is sensitive to electrostatic discharge and may be damaged if
not handled properly. Static charge is generated through the contact of non-
conductive materials such as plastic bags, synthetic clothing and carpeted
floors.
•
It is recommended that a grounded wrist and/or foot band be used when
dismantling the analyzer. If a wrist or foot band is not available, firmly grasp
the metal edge of the RS-232 Serial Port before touching any components.
Repeat often during service.
Warning
•
Before performing any service, remove all Battery Adapters, turn off the
analyzer, and disconnect the power cord.
•
For C7000ER, a high voltage is present on the power supply circuit board
and the capacitors for at least 30 minutes after the power switch is turned off.
To ensure satisfactory discharge of energy, allow the analyzer to sit for at
least 30 minutes before removing the chassis cover.
66
C7000 & 7000ER Battery Analyzer V3.5
Services and Upgrades