13
Replacing the Primary (Input) Fuse
If the analyzer does not power up but the input line is checked and power cord is
firmly inserted, the primary (input) fuse may need to be replaced.
Standard Model (C7000-1, C7000-2, C7000-9)
1.
Remove all batteries and adapters, turn off the analyzer and unplug the power
cord.
2.
Locate the fuse drawer below the power cord socket.
3.
Open the fuse drawer by pulling the release tab outward with a fingernail or
small screwdriver.
4.
If the primary (input) fuse needs to be replaced, remove the fuse from the fuse
drawer and replace the fuse.
5.
Push the release tab back to the fuse drawer until it clicks shut.
6.
Insert the power cord and turn on the analyzer.
Model
Primary Fuse
Size
Replacement
Cadex P/N
C7000-1
T1.6 Amp 250V
5 x 20 mm
Schurter 034.3119
52-546-0160
C7000-2
T1.6 Amp 250V
5 x 20 mm
Schurter 034.3119
52-546-0160
C7000-9
T0.8 Amp 250V
5 x 20 mm
Bussman GDC800mA
52-546-0080
C7000ER
T3.5 Amp 250V
5 x 20 mm
Littlefuse 216 31.5
52-546-0315
Fuse Drawer
67