C7000 & 7000ER Battery Analyzer V3.5
Problem
Possible Reason(s) / Solution(s)
6.
Printing Problems.
Check the serial port setup of the serial printer.
Check that the serial or parallel port is enabled
for correct print device.
Set the other port to Disabled.
The LabelMaker option in the Serial Port
Device or Parallel Port Device of System
Setup is wrongly selected. Select Printer
instead.
Check if the serial cable is correctly configured
(refer to Serial Port Settings for serial cable
pin configurations).
7.
Battery service does not
start.
Ensure that the adapter is firmly inserted into
the analyzer.
Check that the battery is correctly and firmly
inserted into the adapter. Make sure that the
contacts on the battery are meeting the
contacts on the adapter.
Turn on the battery switch if applicable.
Press ENTER after selecting the C-Code and
the service program.
Check that the selected C-Code is correct for
the battery.
8.
Connection between battery
and the analyzer is
intermittent.
The battery may be intrinsically safe. Change
the charge method in the Extended C-Code to
DC Charge; lower the Charge rate and
Discharge rate; reduce the End of Discharge to
0.92V/Cell.