C7000 & 7000ER Battery Analyzer V3.5
Detail
Global
Possible Reason(s) / Solution(s)
Thermistor failure
FAIL 150
Battery or adapter thermistor is open or shorted. Clean the
contacts on the battery. Warm the battery to room
temperature.
The battery thermistor may be damaged. Use another good
battery to check and discard battery if damaged. If the
battery is still good and code 150 (Thermistor failure)
appears, contact Cadex Service.
Trickle charge
TRKL CHRG
The reconditioning process completed and the battery is
being recharged.
The program has specified a trickle charge.
Trickle charge
overheat
FAIL 159
The temperature on the battery has risen over its maximum
setting. The trickle charge stops. The Trickle Charge rate
or the Max. Standby voltage in the Extended C-Code may
have to be reduced. Check that the mAh rating of the
battery matches the C-Code mAh setting. Lower the room
temperature.
For non OEM battery, turn off temperature sensing.
U
User programmed
timeout
END CYCLE
The time programmed in the Custom Program completed.
The program is moving into the next step.