- 67 -
the " " button, the machine body and rotatory bottle will fall down. In the process of falling down,
click the " " button again, to stop falling. Both the rise and fall are equipped with limit devices.
2)
Counterclockwise adjust the lock knob (1) in Figure 3 to remove the
machine head lock state.
3)
Adjust the inclination angle adjustment knob(2), spin the bottle into the appropriate volume
in the bath, and then rotate the lock knob (4) in figure 3 to lock in the
machine head. The
range of the
adjust angle is 15-45°.
Reminder
The range of the inclination angle adjustment of the rotatory bottle shall not exceed the
specified limit, otherwise, it will cause damage to the product.
Rotating speed setting
1)
Turn the rotating speed regulator (2) in Figure 4 slowly clockwise and observe whether the
rotatory bottle rotates smoothly.
2)
Under the condition of ensuring the normal operation of the rotary mechanism, the speed is
slowly adjusted to the target value.
Bath parameter settings
1)
Connect to the power supply, switch on the power.
2)
Press the "SET" key (1) on the "Water bath Control Panel" showed in Figure 6c to enter the
temperature setting Interface. A "SP" character or the current temperature setting value will
be displayed alternately in the temperature display window. Press the " " key to increase the
temperature setting, press the " " key to decrease the temperature setting. Press the "SET"
key again to accept and enter the monitoring mode. The temperature display window will stop
blinking, and the water bath starts to heat, and a” HEAT” character in temperature display
window will flash.
Note: Press " " or " " key once, the temperature setting value increases or decreases by
0.1°C; the temperature setting value will continue changing if the " "or " "being continue
pressed.
Tips: the water bath can be self-calibrated if being found its accuracy of temperature
controlling cannot meet the requirement of your laboratory after a long-time use.