- 24 -
Centrifuge stop: The time required for the centrifuge to drop from the highest speed to 0, it is
recommended not to be less than 1 minute;
Distilling: The time for distilling the solvent of the mixture liquid tank to a low liquid level;
Stop: The overall test delay stop time.
Setup debug
Run the overall process in stages, and each workflow can be tested individually (as shown in Figure
34);
Note: Before manually testing each function, the "centrifuge start" function must be turned on in
advance to prevent the solvent from flowing into the centrifuge chamber and failing to be discharged
in time.
Start Centrifuge: start the centrifuge;
S-chamber fill solvent: Fill the washing chamber (or R rubber asphalt washing chamber) with liquid;
Figure 34
S-motor rotate: Start the stirring motor in the washing chamber (or R rubber asphalt washing
chamber);
Ultrasound start: Start the ultrasonic device at the bottom of the washing chamber (there must be
liquid in the washing chamber);
real-time temperature
Interface selection
Real-time liquid level
display (red is the
detected liquid)
Manual test key for
single function