GasAlertMicroClip
User Manual
Each sensor has a high degree of resistance to common
vapors and gases. A sensor will most likely clear itself if you
move the detector to a clean environment and wait 10 to 30
minutes, but it could take up to a couple of days depending
upon the gas that came in contact with the sensor. Do not
expose a sensor to the vapors of inorganic solvents such as
paint fumes or organic solvents. The
Troubleshooting
section
describes problems caused by a sensor in need of
calibration or replacement.
2.
Allow the detector to charge for 2-3 hours. The
charging indicator flashes on the LCD while the
detector is being charged.
3.
Once charging is complete, the LCD continues to
display
. Remove the adapter and activate the
detector.
Note
If the battery indicator does not display within 5
minutes, refer to the
Troubleshooting
section.
To replace a sensor or sensor filter, refer to the following
figure, table, and set of instructions.
To reach full battery capacity, allow the battery to
fully charge and fully discharge three times.
Charging the detector in temperatures above 113°F
will greatly reduce the number of charges the
detector can accept.
The detector may be hot immediately following
charging. This is normal.
Replacing a Sensor or Sensor Filter
a
Warning
To avoid personal injury, use only sensors
specifically designed for the detector. See the
section
Replacement Parts and Accessories
.
Figure 5. Replacing a Sensor or Sensor Filter
36