GasAlertMicroClip
User Manual
Calibration Procedure
To calibrate the detector perform the following procedure.
Note
Calibrate O
2
in clean air.
If
C
is pressed at any time during the calibration
procedure, calibration is
ABORTED
. The detector
emits a quick beep and the calibration due dates
are displayed before the detector returns to normal
operation.
Start Calibration
Note
Verify that the calibration gas you are using
matches the span concentration value(s) in the
detector. See the section,
Calibration Gas
Concentration
.
1.
In a clean atmosphere, press and hold
C
as the
detector beeps, flashes, and vibrates to the
corresponding
OFF
countdown. Continue to hold
C
as the detector briefly deactivates.
2.
The detector then reactivates and performs the
CAL
countdown. Continue to hold
C
until the
CAL
countdown is complete to enter calibration.
Note
If
C
is not held for the entire
CAL
countdown, the
detector will deactivate.
30