GasAlertMicroClip
Calibration
If you apply gas to a sensor and the detector fails to span the
sensor, repeat the calibration procedure using a new gas
cylinder. If the sensor fails the span a second time, replace
the sensor. See the section,
Replacing a Sensor or Sensor
Filter.
Calibration Due Date
Note
If a sensor does not successfully span, its
calibration due date will not be reset.
6.
After span is complete, the LCD displays the
following calibration due date screens before
returning to normal operation.
Note
The number of days shown in the last screen
represents the earliest date a calibration needs to
be performed.
If a sensor failed to successfully span and it is past its
calibration due date, the LCD displays the following
three screens.
Press
C
to acknowledge the warning before returning
to normal operation.
33