GasAlertMicroClip
Activating the Detector
Press
C
to acknowledge the warning and enter
normal operation, or continue with the self-test if
applicable.
Forced Calibration Enabled
If forced calibration is enabled (refer to
Force Calibration
When Overdue
) and a sensor is past its calibration due date,
calibration is mandatory before the detector can enter normal
operation.
If
C
is pressed, the detector launches the calibration
procedure (refer to
Calibration
). If
C
is not pressed within 2
minutes, the detector deactivates.
Note
Calibration can also be performed by using BW
Technologies by Honeywell’s MicroDock II.
Cal Lock Enabled
If the
Cal Lock
option is enabled, an infrared (IR) device is
required to calibrate the detector.
IR--LOCK
displays as the detector tries to establish a
connection with an IR device (e.g., BW Technologies by
Honeywell’s MicroDock II or IR Link adapter).
Once IR communication has been established, the calibration
procedure is executed. If IR communication is not
established within 2 minutes or
C
is pressed, the detector
deactivates.