Model 9163
Digital indicator
Page 96
If you have now configured all the enabled alarm limits, you are taken to the parameter
Lo
.
AL
, which is
where you specify the lower extreme value for the alarm limit.
PV
¾
Set the lower extreme value.
This value must lie between –19 999 and 99 999.
¾
Press the [F] button briefly to confirm this value.
The display now shows the parameter
Hi
.
AL
.
PV
¾
Specify the upper extreme value here.
This value must also lie between –19 999 and 99 999.
¾
Press the [F] button briefly to confirm the value.
You have now reached the last step of the limit configuration process: parameter
rEL
.
PV
This parameter defines the behavior of the output if a sensor breaks (error code
Err
,
Sbr
and
Ebr
).
¾
Use a code to specify the behavior in the event of a broken sensor.
Code
Limit 1
Limit 2
Limit 3
0 OFF OFF OFF
1 ON OFF OFF
2 OFF ON OFF
3 ON ON OFF
4 OFF OFF ON
5 ON OFF ON
6 OFF ON ON
7 ON ON ON
Add-on function:
•
+16, to set limit 4 to "ON"
¾
Press the [F] button briefly to confirm the input.
Configuration of the alarm limits is now complete and you are back in the main menu.