Model 9163
Digital indicator
Page 76
You use this parameter to set the digital filter for the input concerned.
You can set the filter in a range between 0.00 and 20.00 seconds.
Note:
The digital filter
F1t
is a display filter, i.e. it affects the display.
Note:
Set the digital filter to the value "0" to disable it.
Once you have set the digital filter:
¾
Press the [F] button.
You are now at the parameter
dPS
.
PV
This parameter is used to set the position of the decimal point.
¾
Use a code to set the position.
Only positions "0" and "1" are available for thermocouples.
Code Format
0
0 0 0 0 0
1
0 0 0 0
.
0
2
0 0 0
.
0 0
3
0 0
.
0 0 0
4
0
.
0 0 0 0
Add-on function:
•
+8 disables the message
Lo
and
Hi
.
Once you have set the position of the decimal point for this input:
¾
Continue by setting the upper and lower scale values.
To do this, press the [F] button.
The display now shows the parameter
LoS
.
PV
¾
Set the lower scale value.
Once you have set the lower scale value: