Digital indicator
Model 9163
Page
133
10.5.4 Thermocouple (TC)
These active sensors are made of two wires of different metals or metal alloys fused together at one end.
When the fused junction is heated, a thermal EMF is produced at the wire ends. The amplitude of the
EMF depends on the type of metals used and the temperature difference between the measuring point
and the reference junction.
Materials used nowadays comply with the set of fundamental values given in DIN EN 60584 and
DIN 43710.
Perform the following steps to carry out a calibration:
Note:
Before a calibration, connect all the sensors to the digital indicator. If all the sensors that are connected to
the digital indicator during normal operation are not connected, measurement errors can arise in the
calibration.
Note:
If necessary, you can abort the fine adjustment procedure for the analog output from the parameter
tc
.
00
onwards. To abort the process, hold down the [ ] + [F] buttons simultaneously. You are then taken back
to the parameter
U
.
CAL
.
¾
Connect the sensors to the instrument.
¾
First go to the menu
PAS
.
To do this, hold down the [F] button until
PAS
is displayed.
PV
¾
Use the
▲
or
▼
buttons to set the value "
99
".
Hold down the [F] button until
U
.
CAL
is displayed.
F
PV
¾
Select the relevant input:
Code Function
1 Input
IN1
2 Input
IN2
¾
Press the [F] button briefly to confirm your selection.
Summary of Contents for 9163
Page 4: ...Model 9163 Digital indicator Page 4...
Page 5: ...Digital indicator Model 9163 Page 5 9163 as panel mount unit 9163 as bench top unit...
Page 6: ...Model 9163 Digital indicator Page 6...
Page 8: ...Model 9163 Digital indicator Page 8...
Page 12: ...Model 9163 Digital indicator Page 12...
Page 56: ...Model 9163 Digital indicator Page 56...
Page 68: ...Model 9163 Digital indicator Page 68...