Digital indicator
Model 9163
Page
121
You have selected one of the linearization types 0 to 4:
Once you have selected the linearization type or scale for the temperature sensor, you reach the
parameter
Step
.
n
.
PV
¾
Set the number of intervals (steps) you require.
You must set a value of at least one and no more than 32.
Once you have confirmed the number by pressing the [F] button briefly, you access the parameter
S
.
00
.
PV
¾
Set the lower scale value.
Enter here the same value that you have set for this input in the parameter
LoS
. This value must
lie between –19 999 and 99 999.
After confirming ([F] button), the display shows the parameter
S
.
01A
.
PV
If linearization involves "teach-in", the input is measured. You need to confirm this value by
pressing the [F] button.
¾
Set here the assigned / measured input value in multiples of 1/10 000 of the value that you will
specify in the parameter
S
.
01b
.
The value of the parameter
S
.
01b
must lie between 0 and 10 000.
PV
Repeat these steps for the number of intervals that you have specified in the parameter
Step
.