Digital indicator
Model 9163
Page
69
8. Configuring the inputs and outputs
The advanced Configuration / Parameterization menus for the 9163 digital indicator contain parameters
that let you define the finest points of the instrument configuration. This means that the digital indicator
can satisfy the requirements of practically any application.
Warning!
Warning of personal injury and property damage!
Make sure that all the parameter sets are correct before operating the
instrument.
The following pages provide information on the various menus of the 9163 indicator in turn. Each
parameter is accompanied by details of its function, its default value if any, and the limits within which it
can be set.
Note:
Please refer to the values in the tables when setting parameters. For some parameters you also need to
add on values to obtain certain functions.
8.1 Configuring the inputs
8.1.1 Configuring a main input
You configure the two main inputs of the 9163 digital indicator in the menus
InP
.
1
and
InP
.
2
. Both menus
have an identical structure and the same settings. Each menu contains the settings for one main input.
¾
Switch from the main menu into the relevant configuration menu (
InP
.
1
or
InP
.
2
).
To do this, hold down the [F] button until
InP
.
1
or
InP
.
2
is displayed.
PV
¾
Press the [F] button briefly once.
You are now at the parameter
typ
.
PV
¾
Set a specific sensor type for the input by entering the sensor class.
Summary of Contents for 9163
Page 4: ...Model 9163 Digital indicator Page 4...
Page 5: ...Digital indicator Model 9163 Page 5 9163 as panel mount unit 9163 as bench top unit...
Page 6: ...Model 9163 Digital indicator Page 6...
Page 8: ...Model 9163 Digital indicator Page 8...
Page 12: ...Model 9163 Digital indicator Page 12...
Page 56: ...Model 9163 Digital indicator Page 56...
Page 68: ...Model 9163 Digital indicator Page 68...