Digital indicator
Model 9163
Page
71
Add-on functions:
•
+32 for a sensor-specific linearization.
•
+64 for thermocouples with external compensation element.
Note:
You can use class settings 28 and 29 without calibrating the sensor. Set the required parameters
for offset and sensitivity.
For class settings 28 and 29, a 10 V supply voltage corresponds to maximum sensitivity of
4 mV/V.
Once you have set the sensor class:
¾
Press the [F] button briefly.
The display now shows the parameter
F1t
.
PV
You use this parameter to set the digital filter for the input concerned.
You can set the filter in a range between 0.00 and 20.00 seconds.
Note:
The digital filter
F1t
is a display filter, i.e. it affects the display.
Note:
Set the digital filter to the value "0" to disable it.
Once you have set the digital filter:
¾
Press the [F] button briefly.
You are now at the parameter
dPS
.
PV
This parameter is used to set the position of the decimal point.