Digital indicator
Model 9163
Page
21
3. Electrical connections
Caution!
Risk of electric shock!
The 230 V unit has Class II protection and is classified as
Installation Category II
Instruments with 20...27V AC/DC power supply must only be supplied from a
current source with Class III protection.
The panel-mount unit does not have an On/Off switch!
Connect a two-pole circuit breaker (with CE mark) in the input supply to the
instrument to disconnect the power supply. The circuit breaker must be
installed in the immediate vicinity of the instrument within easy reach of the
user.
One circuit breaker can be used for more than one instrument.
All connecting terminals are on the rear of the unit.
Section 13: "Technical specifications" contains the technical specifications.
Table 1: Cables for electrical connections to digital indicator
Function Cable
type
Length
Connecting lead
up to 1 mm
2
1
m
Connecting leads to relay output
up to 1 mm
2
3.5
m
Serial connection cable
up to 0.35 mm
2
3.5
m
Thermocouple input
up to 0.8 mm
2
compensated
5 m
Input for strain gage sensors, potentiometers, linear
signals from PT100 resistance thermometers
up to 1 mm
2
3
m
Analog outputs for signal feedback
up to 1 mm
2
3.5
m
Digital inputs / outputs
up to 1 mm
2
3.5
m
Summary of Contents for 9163
Page 4: ...Model 9163 Digital indicator Page 4...
Page 5: ...Digital indicator Model 9163 Page 5 9163 as panel mount unit 9163 as bench top unit...
Page 6: ...Model 9163 Digital indicator Page 6...
Page 8: ...Model 9163 Digital indicator Page 8...
Page 12: ...Model 9163 Digital indicator Page 12...
Page 56: ...Model 9163 Digital indicator Page 56...
Page 68: ...Model 9163 Digital indicator Page 68...