3 Transport and storage
Only transport the product inside the original packaging or a suitable alternative.
The equipment must be protected from moisture and heat when not in use. They must be stored in a covered, dry and dust-free
room at a temperature between -20 °C to +40 °C (-4 °F to 104 °F). To avoid bearing damage, ensure a vibration-free environment
(v eff<0.2 mm/s).
Outdoor storage is
prohibited
. On principle the operator must meet all applicable standards with respect to preventing damage
due to lightning, which could result in sample gas pump damage.
Storage areas must not contain any equipment generating ozone, e.g. fluorescent lighting, mercury vapour lamps, high voltage
electrical equipment.
After prolonged storage or downtimes test the insulation resistance of the winding, phase against phase and phase against
mass, prior to initial operation. Moist windings can cause current leaks, flashovers and breakdown. The insulation resistance of
the stator winding must be at least 1.5 MΩ measured at a winding temperature of 20 °C (68 °F). Values below this require drying
the winding.
The motor shaft should be turned occasionally to ensure the entire bearing remains lubricated. To do so, remove the three cross-
tip screws (9) of the console cover (8) and remove. This exposes the crank gear (10). You can now turn the motor shaft on it.
For the item numbers, please refer to the assembly drawing 42/025-Z02-01-2 in the appendix.
CAUTION
Contusion hazard
Contusion of the fingers
Don’t have your fingers caught between eccenter and slide.