©2004,2002 BRUNO INDEPENDENT LIVING AIDS, INC.®
37
SRE-2700 06-11-2004
Check batteries in remote call / send unit.
Check IR receiver.
Check for loose connection.
Check that transmitters are `learned' to the same switch position
(see Learning the Transmitter).
Check limit switch in carriage assembly for proper operation.
Unit does not shut off
when it hits the bumper
at the end of the rail
Unit will not
operate with
call / send
remote transmitter
Unit operates
erratically or
intermittently with a
rider using the
armrest-mounted
control switch
Check to see that the footrest safety tray is not dragging on the
stair nosing or hitting debris on the stairs. Clear debris and, if
necessary, reposition the stair rail mounting brackets to correct
the problem.
This is correct lift operation, a safety switch in the seat swivel
prevents the unit from operating with the seat "out of position".
Unit will not
operate unless
the seat is
positioned so that
it faces the open side of the
stairs
TROUBLESHOOTING