Delivery and Transport
Site Planning Information BioSpec 70/30 USR_2_V020
83
7.4
Transporting Electronic Cabinets
The electronics cabinets contain sensitive and sometimes very heavy assemblies.
Sometimes the cabinets have a high center of gravity and are therefore in danger of tipping.
The cabinets may
not be tipped
during transportation, they may
not
get wet
and they may
not be subject to high acceleration
and
jolting
.
7.5
Moving into the building
Before the MR instrument can be delivered and moved to the installation site, the load-
bearing capacity along the transport route must be calculated and verified by a qualified
structural engineer.
The building dimensions along the entire transport routes must be carefully checked and
must allow the transportation according to the transport requirements given below.
For that purpose, the transport weights of magnet and cabinets are provided in chapter
, the transport dimensions of the magnet are given in the figure
below.
CAUTION
Risk of personal injury during transportation within buildings.
Occupational standards and minimum safety distances between building structures and
heavy loads must be fulfilled along the entire transport route to avoid squeezing of hands or
feet.
u
Only trained, qualified, and authorized personnel must transport the MR instrument on
the transport routes from unloading to final destination within the building.
Summary of Contents for BioSpec 70/30 USR V2
Page 6: ...Contents vi Site Planning Information BioSpec 70 30 USR_2_V020...
Page 20: ...Safety 20 Site Planning Information BioSpec 70 30 USR_2_V020...
Page 88: ...Moving or Dismantling the MR Instrument 88 Site Planning Information BioSpec 70 30 USR_2_V020...
Page 98: ...Tables 98 Site Planning Information BioSpec 70 30 USR_2_V020...
Page 103: ...Site Planning Information BioSpec 70 30 USR_2_V020 103...