The galvanic insulation must be installed in each case to avoid ground loops or RF
couplings.
Figure 6.8: Drawing of the magnet exhaust outlet (1) and the exhaust system (2). Note that (1) is
supplied by Bruker and (2) is to be installed by the System Owner.
Position of the connection point between the magnet exhaust outlet and the
exhaust system for 70/30 USR V2
Q
Distance of the magnet exhaust outlet from the
center of the magnet dewar.
37.7 cm
D
Diameter of the magnet exhaust outlet at the
connection point.
8.4 cm
H
Height where the exhaust system provided by
System Owner has to end. Please note the
minimum height and maximum tolerance (see also
drawing) to ensure appropriate connection to the
magnet exhaust outlet.
232 + 1.2 cm