Planning Details
Site Planning Information BioSpec 70/30 USR_2_V020
73
Figure 6.21: Relation between the Primary Water Circuit inlet temperature and the required Primary Flow
for the maximum cooling requirement of 12 kW (standard configuration without CryoProbe option).
Water quality
Specification
pH range
6.5 – 8.2
hardness
5.6 – 11.2 °d
maximum proportion of glycol in the water
30 %
chloride
less than 100 mg/l
solid floating particles
less than 10 mg/l
maximum particle size
less than 400 µm
Table 6.20:
Specifications for the water quality